26/09/2019
- 01:36 PM INCLINE - ICP Bug #3424 (Resolved): Mains water supply interruption
- 11am: switched pump controller on/off.
Allowed controller to heat up to ~ 60 °C.
Switched on. Turbo OK.
1:30pm Ba... - 10:11 AM INCLINE - ICP Bug #3424 (Resolved): Mains water supply interruption
- The mains water supply was interrupted during the morning (6 am ish).
Primary pump overheated, so shut down.
Rise...