Project

General

Profile

Actions

SPARE Calibrations » History » Revision 6

« Previous | Revision 6/10 (diff) | Next »
Alexis Jeandet, 15/05/2016 07:05 PM


SPARE Calibrations

Transfer Function

1) Transfer Functions @F0

For the whole test we will use the following parameters:
  • LFR mode= Normal Mode
  • ASM period = 4s
  • Snapshots period = 22s
  • Input signal amplitude 1.4Vp

a) Low Frequencies and FFT frequencies

7 frequencies: from 12Hz to 84Hz with ∆f = 12Hz
128 frequencies: from 96Hz to 12288Hz with ∆f = 96Hz

View measurements...

b) Frequencies > F0/2

24 frequencies: from 13056Hz to 30720Hz with ∆f = 768Hz

View measurements...

2) Transfer Functions @F1

a) Low Frequencies

  • LFR mode= SBM1
  • ASM period = 32s
  • Snapshots period = 32s
  • Input signal amplitude 3Vp
  • Step duration = 420s

10 frequencies: from 0.01Hz to 0.1Hz with ∆f = 0.01Hz

View measurements...



Updated by Alexis Jeandet about 8 years ago · 6 revisions

Also available in: PDF HTML TXT