Project

General

Profile

Actions

Bug #175

closed

PA_LFR_ACQUISITION_TIME=0x000000000000 sur TM_LFR_SCIENCE_NORMAL_CWF_F3 incoherent / TC_LFR_ENTER_MODE

Added by Veronique bouzid about 10 years ago. Updated about 10 years ago.

Status:
Closed
Priority:
Normal
Category:
-
Target version:
-
Start date:
19/06/2014
Due date:
% Done:

0%

Estimated time:
revision:
r0

Description

On demande un chgt de mode instantané

08:21:11.241669, TC_LFR_ENTER_MODE (CP_LFR_MODE=1), CP_LFR_ENTER_MODE_TIME=0x000000000000
08:21:11.261724, TM_LFR_TC_EXE_SUCCESS, TIME=0x80000004e88b

ensuite arrive une TM_LFR_SCIENCE_NORMAL_CWF_F3
08:26:47.238015, TM_LFR_SCIENCE_NORMAL_CWF_F3, PA_LFR_ACQUISITION_TIME=0x000000000000,
PA_LFR_CWF3_BLK_NR=672

Question:
Pourquoi PA_LFR_ACQUISITION_TIME=0x000000000000 alors que le chgt de mode n'a été fait qu'à TIME=0x80000004e88b
soit 4secondes apres le début du reset HW.

Dans ce script, on n envoie pas de Timecode.

-------------------------------------------------------
Contexte:
LPPMON: Version=0.2.2 Branch=default Changeset=835955994d5f
Carte mini-LFR: LFR-172200 dev V1.0; No série III (sans connecteurs sub-click)
Vhdl: mini-lfr_0.1.23
Brique Star-Dundee S/N <illisible>.
Soft:1.0.0.11 (variante sur carte finale)

TEST CASE = SVS-0073
Req = SSS-CP-FS-431

Actions #1

Updated by paul leroy about 10 years ago

  • Status changed from New to Resolved
  • Assignee changed from paul leroy to Veronique bouzid

fsw >= 1.0.0.12

Bug identifié et corrigé. Vérification sur deux réception de CWF_F3 consécutives.

Actions #2

Updated by Veronique bouzid about 10 years ago

  • Status changed from Resolved to Closed

Test (SVS-0073, script loadNorm.py) rejoué en 1.0.0.12

C'est Ok.
16:46:51.727297, TC_LFR_ENTER_MODE (CP_LFR_MODE=1), CP_LFR_ENTER_MODE_TIME=0x000000000000
16:46:51.78917, TM_LFR_TC_EXE_SUCCESS,*TIME=0x80000004c5a8*

16:49:39.790752, TM_LFR_SCIENCE_NORMAL_CWF_F3, CCSDS_VERSION_NUMBER = 0, PACKET_TYPE: TM_PACKET = 0, DATA_FIELD_HEADER_FLAG: WITH_HEADER = 1, PROCESS_ID: RPW_PID_2 = 76, PACKET_CATEGORY: PRIVATE_SCIENCE_OR_TELECOMMAND = 12, (PACKET_ID=0xccc), SEGMENTATION_GROUPING_FLAG: STANDALONE_PACKET = 3, SEQUENCE_CNT=147, (PACKET_SEQUENCE_CONTROL=0xc093), PACKET_LENGTH=4051, SPARE_1=0, PUS_VERSION = 1, SPARE_2=0, SERVICE_TYPE: SCIENCE_DATA_TRANSFER = 21, SERVICE_SUBTYPE: SCIENCE_REPORT = 3, DESTINATION_ID: GROUND = 0, TIME=0x80000004d0f0, PA_LFR_SID_PKT: SC_N_CWF_F3 = 1, PA_BIA_MODE_MUX_SET: SET_0 = 0, PA_BIA_MODE_HV_ENABLED: ENABLED = 1, PA_BIA_MODE_BIAS1_ENABLED: ENABLED = 1, PA_BIA_MODE_BIAS2_ENABLED: ENABLED = 1, PA_BIA_MODE_BIAS3_ENABLED: ENABLED = 1, PA_BIA_ON_OFF: OFF = 0, PA_LFR_ACQUISITION_TIME=0x80000004d0f0, PA_LFR_CWF3_BLK_NR=672

-----------------------------------------------------------
Contexte:
SocExplorerEngine.getSocExplorer: Version = 0.2.2, Branch = default, Changeset = c839740ef520
Carte EM
Vhdl: EM_1.1.23
Brique Star-Dundee S/N <illisible>.
Soft:1.0.0.12 (variante sur carte finale)

TEST-CASE = SVS-0073

Actions