INCLINE - ICP: Issueshttps://hephaistos.lpp.polytechnique.fr/redmine/https://hephaistos.lpp.polytechnique.fr/redmine/redmine/favicon.ico?15080976012019-09-26T08:11:23ZRedmine
Redmine Bug #3424 (Resolved): Mains water supply interruptionhttps://hephaistos.lpp.polytechnique.fr/redmine/issues/34242019-09-26T08:11:23ZGarrett Curley
<p>The mains water supply was interrupted during the morning (6 am ish). <br />Primary pump overheated, so shut down. <br />Rise in backing pressure caused turbo pump driver overload and shut down.<br />Chamber rose to atm pressure.<br />Both Baratrons out of range.<br />10:00am: Primary pump started. OK<br />1 Torr Baratron OK.<br />0.1 Torr Baratron could have issue. Switched off just in case. <br />10:10am; Turbo pump restart attempted. NOT OK. TMS low temp error.</p> Task #3421 (New): Repair Cl2 mass flow controllerhttps://hephaistos.lpp.polytechnique.fr/redmine/issues/34212019-09-19T09:40:16ZGarrett Curley
<p>Defective Brooks Model 5850TR</p>